HAMILTON-MR1 Specifications

Dimensions

  • Height: 10.43” (26.5 cm)
  • Width: 11.1” (30 cm)
  • Length: 8.3” (21 cm)
  • Weight: 15 lb (6.8 kg)

Pneumatics

  • High-pressure Oxygen Inlet
    • Pressure: 2.8 to 6 bar / 280 to 600 kPa / 41 to 87 psi
    • Flow: Maximum of 200 l/min
    • Connector: DISS (CGA 1240) or NIST
  • Air Supply: Integrated blower
  • Delivered flow
    • 260 l/min ±10% against ambient pressure (at sea level); 120 l/min at 30 cmH20; 0 to 200 l/min with 100% O2; Flow limitation in neonatal modes: 40 l/min
  • Flow accuracy (for calibrated flow sensor)
    • Adult/Ped: ±10% or ±300 ml/min (whichever is greater)
    • Neonatal: ±2 ml/s or ±10% (whichever is greater)
  • Delivered pressure
    • Adult: 0 to 60 cmH2O
    • Neonatal: 0 to 45 cmH20
  • Inspiratory outlet (To patient port) Connector: ISO 15 mm female/22 mm male conical
  • Expiratory outlet (From patient port) Connector (on the expiratory valve): ISO 15 mm female/22 mm male conical

Battery

  • Electrical: 10.8 V DC, 6.7 Ah, 72 Wh, 50 W typical, 150 W maximum
  • Type: Lithium-ion
  • Operating time: Operating times are measured with two fully charged batteries. Maximum run time: 9 hours.
  • Recharge time: While the ventilator is connected to primary power, approximately 3.25 h to fully recharge one battery, approximately 6.25 h to fully recharge two batteries.

Magnetic Field

  • Static magnetic field
    • ≤ 50 mT
    • Corresponds to approximately 1 m distance from the front of a 3.0 T MRI scanner.

Electrical

  • Input Power: 100 to 240 VAC -15% /+10%, 50/60 Hz
  • Power Consumption: 50 W typical, 120 W maximum
Category:

HAMILTON-MR1 Features

The HAMILTON-MR1 MRI ventilator provides positive pressure support for adult and pediatric patients. This Ventilator from Hamilton Medical can be used in an MRI department and is a good fit for patient transport within a medical facility. This is a fully functional electronically controlled ICU ventilator. The MR1 ventilator can be used in a magnetic field strength of 50 mT.

  • Adult, pediatric, and optional neonatal ventilation compatibility.
  • High-performance turbine allows this ventilator to run independently from compressed air.
  • Leak compensation for NIV and invasive ventilation.
  • Up to 9 hours runtime on battery power.

Ventilation Modes

  • Intelligent ventilation modes with Adaptive Support Ventilation (ASV)
  • Pressure-controlled modes (including biphasic modes)
  • Volume-controlled modes (adaptive)
  • Pressure support modes
  • Modes for noninvasive ventilation

There are no reviews yet.